Operating atmospheric
pressure
525 ~ 1200hPa(525 hPa corresponding to an altitude of
4572m/15000feet)
Storage and transportation
pressure
Withstands Repeated 1.8m drops on a hard surface.
Complies with requirements for road ambulances according to
prEN1789 (clause 6.4) and requirements for shock and vibration
according to ISO 80601-2-55 (clause 201.15.3.5.101, transport).
IRMA CO
2
:
4.5-5.5VDC, Max 1.0W
(power on surge @ 5V less than 350mA during 200ms)
IRMA AX+:
4.5-5.5VDC, Max1.4W
(power on surge @ 5V less than 350mA during 200ms)
Surface temperature (at
ambient temperature
23℃)
IRMA CO
2
:
Max 39 degree C/102FIRMA AX+:
Max 46 degree C/115F
Modified RS-232 serial interface operating at 9600bps
Disposable adult/child:
Adds less than 6mL dead space;
Pressure drop less than 0.3cmH
2
O @ 30LPM.
Disposable infant:
Adds less than 1mL dead space;
Pressure drop less than 1.3cmH
2
O @ 10LPM.
NOTE 1: After being in a condensing atmosphere, this module shall be stored for more
than 24 hours in an environment equivalent to operating humidity.
The humidity range 50-100% is valid for the temperature range -40℃ - 40℃.
Table 9-21 Data Output
The newest breathing value (F
i
and ET) is continually displayed after one
breath.
Gas monitor module CO
2
Automatic agent
identification
IRMA AX+: Primary and secondary agent.