9.Technical data
157
Table 9-23
Accuracy specifications - standard conditionsThe following accuracy specification is
valid for dry single gas at 22+/-5 degree C and 1013+/-40hPa.
Range
1)
Gas
CO
2
AX+
Accuracy
CO
2
015
1525
015
1525
± (0.2vol%+2% of reading)
Unspecified
N
2
O
--
0100
± (2vol%+2% of reading)
HAL,ISO,ENF
--
08
812
± (0.15vol%+5% of reading)
SEV
--
010
1015
± (0.15vol%+5% of reading)
DES
--
022
2225
± (0.15vol%+5% of reading)
NOTE 1: Gas concentration reported in units of volume percent.
Table 9-24
Accuracy specifications –all conditions
1)
Gas
Accuracy
CO
2
± (0.3vol%+4% of reading)
N
2
O
± (2vol%+5% of reading)
Agents
2)
± (0.2vol%+10% of reading)
NOTE 1: The accuracy specification is valid for the operating temperature and humidity
conditions specified, except for interference specified in the table “Interfering gas and
vapor effects” below and the section "effect from water vapor partial pressure on gas
reading". The effect from water vapor partial pressure on gas reading.
NOTE 2: The accuracy specification for IRMA AX+ is not valid if more than two agents
are present in the gas mixture.