2.Components
17
The functional description of components is as follows.
NO
Item
Description
Figure
2
Auxiliary oxygen
outlet
Adjust the knob of the auxiliary oxygen
flowmeter, this port can output oxygen
only with s flow rate of 015 L/min.
3
Auxiliary O
2
output
port
When the range of inlet gas pressure is
280kPa—600kPa, the flow of drive gas
output is more than 90L/min.
WARNING: The requirement flow
of the device connected to the
auxiliary drive gas outlet interface
cannot exceed 90L/min.
5
Common Gas
OutletCGO
It functions as the linkage between the
mainframe and breathing system and
delivers the fresh gas to breathing
system.
8
Drive gas interface
There are 3 interfaces: integrating
sampling tube port, heating gas cable
interface, and drive gas interface. By
connection of 3 interfaces, the breathing
system can connect with the main frame.