Vent:
At the left side of the vent menu, there is a
patient type icon, which can be either adult
or child, and will match the patient icon at the
top of the interface.
In the middle of the vent menu, the low and
high alarm limits of MV, Pressure, and
FreqMIN can be set.
At the right side of the vent menu, the alarm
volume can be adjusted.
Gas:
The low and high alarm limits of FiO
2
, etCO
2
and FiCO
2
can be set.
Agent:
Alarm limits for Sevoflurane (SEV),
Desflurane (DES), Halothane (HAL),
Isoflurane (ISO), and Enflurane (ENF) can
be set.
Log:
The alarm time and details can be viewed
from the log menu. Click on the alarm
message bar the detail information of the
corresponding alarm appears on the screen,
including setting parameters (setting mode,
patient type and parameter), detection
parameters and alarm limits.
Record the number of shutdown in the
system menu, and record the time point of
the shutdown in the configuration menu.