8-24 Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
If You Have A Problem
If Errors Occur
108 Duplicate variable names exist.
109 The setup is not finished.
110 Standby chan cannot use R-BOX resistor
For standby channel, you cannot use R-Box resistor.
111 Common chan cannot use R-BOX resistor.
For common mode channel, you cannot use R-Box resistor.
112 VAR1 step number is out of range.
113 START and STOP have different sign.
115 SMU pulse Period must be >= Width+4ms.
116 VAR1 output power too large for unit.
117 VAR1' output power too large for unit.
118 VAR2 output power too large for unit.
119 TOT SMP TM=AUTO is for LINEAR only.
AUTO can be specified for total sampling time only when LINEAR
sampling mode is selected.
120 Only LINEAR when init int <= 480 us.
When initial interval is set to 480 ms or less, you cannot specify LOG or
THINNED-OUT sampling mode.
121 For LINEAR set AUTO if init int<=480us
When initial interval is set to 480 ms or less and when LINEAR
sampling mode is set, AUTO must be set in TOTAL SAMP. TIME
field.
122 TOT SP TM must be>=INIT INT(NOofSMP-1)
Total sampling time must be set in the following range:
total sampling time ³ initial interval ´ (number of samples -1)
123 STOP CONDITION NAME is not set.
124 PGU pulse Period must be > Width.
125 PGU pulse Period must be >= Delay.