Agilent 4155B/4156B User’s Guide Vol.2, Edition 5 8-23
If You Have A Problem
If Errors Occur
61 ADC time out.
The AD converter has caused a time out. Perform 109: ADC test on the
SYSTEM: SELF-CALIBRATION/DIAGNOSTICS screen. If this test
fails, the 4155B/4156B may need service. Contact the nearest Agilent
Technologies Sales and Service office.
62 ADC FIFO overflow.
The AD converter has caused a FIFO overflow. A data transfer error
occurred between the AD converter and the SMUC. The 4155B/4156B
may need service. Contact the nearest Agilent Technologies Sales and
Service office.
63 SMUC failed to send data to HOSTC.
The SMU controller failed to send data to the host controller. Retry the
measurement. If this error is still displayed, recycle the power by
turning the instrument off and then on again.
64 TIFF format supports SCREEN DUMP only.
The TIFF format is only supported by the SCREEN DUMP function.
65 TIFF image can only be written to a FILE.
A TIFF image can only be written to a FILE. The 4155B/4156B cannot
output TIFF format images to a printer or plotter.
66 HR TIFF format supports DUMP and GRAPH PLOT.
A high resolution TIFF image is only supported by the DUMP and
GRAPH PLOT functions.
67 HR TIFF image can only be written to a FILE.
A high resolution TIFF image can only be written to a FILE.
100 VAR1 is not assigned.
101 VAR1 assigned to multiple Channels.
102 VAR2 assigned to multiple Channels
103 VAR1' assigned to multiple Channels.
104 VAR1 and VAR1' must be same MODE.
105 Cannot set multiple SMUs to pulse mode
106 Cannot use VAR when SAMPLING.
107 Cannot use SMU pulse when SAMPLING.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4155B and is the answer not in the manual?

Agilent Technologies 4155B Specifications

General IconGeneral
BrandAgilent Technologies
Model4155B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals