Agilent 4155B/4156B Users Guide Vol.2, Edition 5 2-17
Measurement Mode
Sampling Measurement Mode
Sampling measurement is executed as explained below:
1. Forces constant current, constant voltage, or pulsed constant bias.
2. Waits hold time.
3. Triggers one point measurement.
4. Measurement unit executes measurement. Measurement result data is stored in
memory.
5. Triggers one point measurement. Interval of trigger is same as Sampling
Interval.
6. (Case 1) Measurement unit executes measurement if it is ready to measure.
Measurement result data is stored in memory.
(Case 2) Measurement unit waits next trigger if it is busy or in measurement.
7. Repeats steps 5 and 6 until that a sampling completion condition is satisfied.
In Figure 2-6, sampling measurement stops when the completion condition total
sampling time is satisfied.
Number of measurement data stored in memory depends on the sampling
completion condition. Maximum number is specified by the NO. OF SAMPLES
field of the MEASURE: SAMPLING SETUP screen. However the measurement
will be immediately stopped if a sampling completion condition is satisfied before
reaching the maximum number. For the sampling completion condition, see
Sampling Completion on page 2-20.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4155B and is the answer not in the manual?

Agilent Technologies 4155B Specifications

General IconGeneral
BrandAgilent Technologies
Model4155B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals