EasyManuals Logo

Agilent Technologies 8163B User Manual

Agilent Technologies 8163B
466 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #419 background imageLoading...
Page #419 background image
Optical Switch modules Error Messages
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition 419
error message KeyDisable(): ERROR
KeyDisable(): disabledCnt >= disabledMax
KeyDisableTask(): received wrong channel number.
KeyDisableTask(): disabledCnt <0.
KeyDisableTask(): couldn’t send myself a msg.
KeyEnablePortF(): wrong interrupt number
KeyDisablePortF(): wrong interrupt number
description Problem when enabling or disabling a front panel key. Probably a
calibration data or firmware problem.
test value
action Please return the unit to your local service center for repair
error message KeySelftest() called without calling KeyInit() before
KeySelftest(): KeyDisable() failed
KeySelftest(): selftestPre() failed
KeySelftest(): key pressed during selftest or hardware problem
KeySelftest(): selftestPost() failed
KeySelftest(): KeyEnable() failed
KeySelftestExecPortF(): wrong interrupt number
KeySelftestPreTPU(): FreeTpultc() failed
KeySelftestExecTPU(): GetTpuDioState() failed
KeySelftestPostTPU(): AllocTpultc() failed
Selftest (stAll, 0) failed
description Selftest problem. Possible causes:
- hardware failure
- key pressed during selftest
- calibration data or firmware incorrect
test value
action Please return the unit to your local service center for repair
All manuals and user guides at all-guides.com

Table of Contents

Other manuals for Agilent Technologies 8163B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8163B and is the answer not in the manual?

Agilent Technologies 8163B Specifications

General IconGeneral
BrandAgilent Technologies
Model8163B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals