Error Messages Optical Switch modules
420 Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
error message InitKey(): unsupported keyLocation
KeyAck(): unsupported keyLocation
KeyEnable(): unsupported keyLocation
KeyDisable(): unsupported keyLocation
KeySelftestPre(): unsupported keyLocation
KeySelftestExec(): unsupported keyLocation
KeySelftestPost(): unsupported keyLocation
description Calibtation data incorrect
test value
action Please return the unit to your local service center for repair
error message Preset (0) failed
Preset(): channel X, config Y failed
description Preset failed, probably due to incorrect calibration data
test value X module channel
Y switch configuration number
action Please return the unit to your local service center for repair
error message Error checking firmware CRC!
description The firmware CRC is wrong
test value
action Please return the unit to your local service center for repair
error message Unknown internal error during init
description Error with unknown cause occured
test value
action Please return the unit to your local service center for repair
error message InitDefaultCaldata(): couldn’t alloc defaultCalData
InitDefaultCaldata(): only 2 channels supported in default caldata
InitDefaultCaldata(): unknown type
description This only occurs when running special debug software
test value
action Please return the unit to your local service center for repair