EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>89410A

Agilent Technologies 89410A User Manual

Agilent Technologies 89410A
339 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #245 background imageLoading...
Page #245 background image
Manual sweep
For swept spectrum measurements, sweep times can be very long for low
resolution bandwidth to span ratios. You can use manual sweep to disable the
analyzer’s automatic sweep. Manual sweep lets you tune the analyzer to a discrete
frequency. This lets you measure the amplitude of discrete frequencies without
waiting for the analyzer to sweep through an entire span—a considerable
advantage when using a narrow resolution bandwidth.
Manual sweep is also useful when making automated measurements. Using manual
sweep dramatically reduces measurement time since it’s much faster to transfer
amplitude data for a single frequency over the GPIB rather than sending data for
the entire display.
Fundamental Measurement Interactions
19 - 13

Table of Contents

Other manuals for Agilent Technologies 89410A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 89410A and is the answer not in the manual?

Agilent Technologies 89410A Specifications

General IconGeneral
BrandAgilent Technologies
Model89410A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals