7. Load the modified trace-buffer file into one of the analyzer’s data registers.
On your computer, copy the modified trace-buffer file to floppy disk.
Insert the disk in the analyzer’s disk drive.
Press [
Save/Recall], [recall more], [catalog on] and select your file.
Press [
recall trace buffer], [from file into D1].
8. Configure the arbitrary source to use the data register.
Press [
Source
], [
source type
], [
arb data reg], [D1
]
,[
Return
].
9. Turn on the arbitrary source.
Press [
arbitrary], [Return], [source on].
The analyzer stores trace data in Standard Data Format (SDF). You
substitute trace data from a waterfall or spectrogram display using the
sdfydata Standard Data Format (SDF) utility. This utility is one of several
utilities included in the Standard Data Format Utilities: User’s Guide
shipped with your analyzer. The sdfydata utility automatically converts the
ASCII data to SDF format as it copies the ASCII data into the SDF file.
For additional details about installing and using the SDF utilities, see the
Standard Data Format Utilities: User’s Guide shipped with your analyzer.