352 6000 Series Oscilloscope User’s Guide
9 Characteristics and Specifications
Sensitivity For ± 1V range setting:
DC to 100 MHz, 100 mV;
MSO/DSO6032A/6052A/6102A
>100 MHz to bandwidth of oscilloscope: 200 mV
For ± 8V range setting:
DC to 100 MHz, 250 mV;
MSO/DSO6032A/6052A/6102A
>100 MHz to bandwidth of oscilloscope: 500 mV
MSO/DSO6014A:
DC to 100 MHz: 500 mV
MSO/DSO6034A/6054A/
6104A:
DC to 500 MHz: 500 mV
Coupling AC (~3.5 Hz), DC, noise reject, HF reject and
LF reject (~50 kHz)
Probe ID MSO/DSO601xA: Auto probe sense
MSO/DSO603xA/605xA/610xA: Auto probe sense and
AutoProbe interface
Agilent- and Tektronix-compatible passive probe sense
Display system
Display 6.3-inch (161 mm) diagonal color TFT LCD
Throughput of oscilloscope
channels
Up to 100,000 waveforms/sec in real-time mode
Resolution XGA –
768 vertical by 1024 horizontal points (screen area);
640 vertical by 1000 horizontal points (waveform area)
256 levels of intensity scale
Controls Waveform intensity on front panel. Vectors on/off; infinite persistence on/off,
8 x 10 grid with intensity control
Built-in help system Key-specific help (in English) displayed by pressing and holding key or softkey of interest
Real-time clock Time and date (user adjustable)
Measurement features
Automatic measurements Measurements are continuously updated. Cursors track last selected measurement.
Voltage (scope channels only) Peak-to-peak, maximum, minimum, average, amplitude, top, base, overshoot, preshoot,
RMS, standard deviation
Time Frequency, period, + width, – width and duty cycle on any channel.
Rise time, fall time, X at max Y (time at max volts), X at min Y (time at min volts), delay,
and phase on oscilloscope channels only.
Counter Built-in 5-digit frequency counter on any channel. Counts up to the oscilloscope’s
bandwidth (1 GHz max). The counter resolution can be increased to 8 digits with an
external 10 MHz reference.
Threshold definition Variable by percent and absolute value; 10%, 50%, 90% default for time measurements

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies MSO6012A and is the answer not in the manual?

Agilent Technologies MSO6012A Specifications

General IconGeneral
BrandAgilent Technologies
ModelMSO6012A
CategoryTest Equipment
LanguageEnglish

Related product manuals