Depending on which combination of Acquisition, Readout & Triggering modes is selected, various timing
parameters become available as follow:
• Exposure Time (secs)
• No. of Accumulations
• Accumulation Cycle Time (secs)
• Kinetic Series Length
• Kinetic Cycle Time (secs)
When acquiring data small changes in the ambient temperature and/or in the heat generation of the driving
electronics within the camera may cause some drift in the baseline level. This is most often observed during
long kinetic series.
Any drift in the baseline level can be corrected by using the Baseline Clamp option. Baseline Clamp corrects
each individual image for any baseline drift by subtracting an average bias signal from each image pixel and
then adding 100 counts to ensure that the displayed signal level is always a positive number of counts.
NOTE: Do not use this function with 5MHz readout (focus mode)