Chapter 8 — Vector Voltmeter, Option 15 8-4 Simple Measurement Using CW Display
MS20xxA VNA Master UG 10580-00166 Rev. C 8-5
11. Connect the DUT for measurement. The display should appear similar to the image in
Figure 8-4.
12. Specify
Measurement Type to be Return for a 1-port DUT. After a 2-port calibration, both
Return and Insertion measurement types can be viewed.
13. Choose to view the measurement results for the return measurement as dB, VSWR, or
Impedance by pressing the
Return Meas. Format soft key. For insertion measurements,
only the dB selection is available.
This completes the procedure for performing a simple measurement using the vector
voltmeter mode.
Figure 8-4. Continuous Wave Menu