EasyManua.ls Logo

APS CPS100 Series - Page 209

APS CPS100 Series
231 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CPS100 SERIES OPERATION MANUAL
SECTION 13:
Error Messages & Troubleshooting
Entire Contents Copyright 2018 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
APS CPS100 Series Power Source Operation Manual Page 209 of 231 PN 160953-10
Move the cursor to the Vdc offset field.
Enter the DMM Vdc reading using the touch
keypad or [0] ~ [9] keys, then press the
[Enter] key to confirm.
Repeat this step until the DMM reading is
between -10 mV and +10 mV.
Wait 2 seconds, then press the Continue soft
key to proceed to the next calibration point.
Put the DMM in mode.
Move the cursor to the Vac = field and enter
the Vac reading using the touch keypad or [0]
~ [9] keys, then press the [Enter] key to
confirm.
Follow the on-screen prompts to complete calibration the 150V AC range. Both AC & DC
output and AC & DC measurements are calibrated. This will require changing the DMM
measurement functions as prompted.
Once completed, repeat the same process for the 300V AC range.

Table of Contents