BD FocalPoint™ GS Imaging System Instrument User’s Manual
130
Figure 13-3 – Slide thickness, adhesive and coverslip mounting specifications
13.2.5 Staining and Fixative Requirements
The slides must be stained with the Papanicoloau method.
13.2.6 Barcode Formats
The barcode types supported by the BD FocalPoint™ Slide Profiler are listed below.
Interleaved 2-of-5
Code 128 A, B, and C
Code 39 and Code 39 Full ASCII
EAN 128
DataMatrix
PDF417
Aztec compact
Barcode labels must be durable so that they remain both human and machine readable under
normal laboratory handling and application to microscope slides. Slides may be exposed to and
immersed in the following solvents:
Water
Ethanol
Xylene
Hematoxylin in 4% Acetic Acid
0.5% HCL
1.5% NH
4
OH in 70% Alcohol
Eosin-Alcohol (EA) stains
EA-36
EA-50
EA-65
Isopropyl Alcohol
For additional questions about barcode label specifications contact BD Technical Service and
Support.
Specimen
Microscope Slide
PP2YHUDOO7KLFNQHVVPP
&RYHUVOLS$GKHVLYH
%DUFRGH/DEHO
6OLGH7UD\
7UD\5HIHUHQFH6XUIDFH
6OLGH6XSSRUW3DG
PPDGKHVLYHDQGJODVVFRYHUVOLSPP
PPDGKHVLYHDQGSODVWLFFRYHUVOLSPP
6SHFLPHQ
0LFURVFRSH6OLGH