28 System Requirements
28.1 BD FocalPoint™ GS Review Station
The BD FocalPoint™ GS Review Station hardware requires a minimum 15 Amp, 50/60 Hz circuit
with proper grounding, and must meet local requirements.
28.1.1 Electrical Specifications
Stage Controller
Operating Voltage 110 volts 230 volts
Operating Frequency 50/60 Hz 50/60 Hz
Current 0.35 Amps 0.18 Amps
NOTE
A range of microscopes are compatible with the BD FocalPoint™
GS Review Station Autostage. Consult the microscope
manufacturer's manual for electrical specifications. Common
applicable microscopes include:
Microscope
Operating Voltage 100–240 VAC
Operating Frequency 50/60 Hz
Current: Nikon 55i 0.60–1.20 Amps
Current: Nikon Eclipse
Ci-L
0.185–0.37 Amps
Current: Olympus
BX43
0.20–0.40 Amps
NOTE
Ratings are dependent on the configuration of the type of
microscope used, light source and power adapter used. Ratings
above are valid for the manufacturers respective LED light source
options and supplied power adapters.