BD FocalPoint™ GS Imaging System Instrument User’s Manual
180
5 Ranking information: All successfully processed slides having an action of Review or QC
Review receive a group (or quintile) ranking, an individual ranking based on their scoring
position within the print set, and their individual rank expressed as a percentage.
For the group ranking, slides are assigned to a numerical quintile between 1 and 5. Within
the print set, the 1st quintile contains the 20% of slides most likely to contain abnormalities,
the 2nd quintile contains the 20% of slides next most likely to contain abnormalities, and so
on, with the 5th quintile containing the 20% of the slides least likely to contain abnormalities.
Each slide receives an individual rank from 1 to n, where a rank of 1 indicates a slide most
likely to contain abnormality and n is the slide least likely to contain abnormality (n is the
number of Review slides in a print set, minus the scant cellularity slides).
For example, the first slide listed on this sample report received an individual rank of 27/41,
or 27 out of 41. For more information on ranking see Section 10.4.1.
The percentage value is the individual rank expressed as a percentage of n.
6 Action required:
Review/Process Review/QC Review/Rerun/Empty. (See Section 15.7).
7 The specimen results for each slide:
Specimen adequacy
Squamous cells. Detected/Insufficient
Endocervical cells. Detected/Insufficient
Inflammation/Obscuration. Less than 50% (Satisfactory), between 50–75% (SBLB), or
greater than 75% (Unsatisfactory).
8 Date: The date and time each slide completed processing.
9 Process Status: A message appears in this area indicating the status of the scan. The
message most likely to be seen is
Scan Completed. However, when slides are rejected, or
the scanning process did not complete, a reason is printed under the Message column of the
report.
10 Additional information is printed as appropriate in the message column. Refer to Section 21
for a description of these messages.