Programmable DC Power Supply (with Solar Array Simulation) 62000H Series
Operating & Programming Manual
*IDN? Identification Query
Type: System interface
Description: This query requests the 62000H to identify itself.
Query Syntax: *IDN?
Query Example: *IDN?
String Description
CHROMA ATE Manufacturer
62150H-600S Model name
123456 Serial No.
01.00 Firmware version
Return Example: CHROMA ATE, 62150H-600S, 123456, 01.00
*OPC Operation Complete Command
Type: Device status
Description: This command causes the interface to set the OPC bit (bit 0) of the
Standard Event Status register when the DC Power Supply has
completed all pending operations.
Syntax: *OPC
Parameter: None
*OPC? Operation Complete Query
Type: Device status
Description: This query returns an ASCII “1” when all pending operations are
completed.
Query Syntax: *OPC?
Return Parameter: <NR1>
Query Example: 1
*RCL Recall Instrument State Command
Type: Device status
Description: This command restores the High Slew Rate Load to the state that was
previously stored in memory with the *SAV command to the specified
location (see *SAV).
Syntax: *RCL <NR1>
Parameter: None
Example: *RCL 1
*RST Reset Command
Type: Device status
Description: Reset System
Syntax: *RST
Parameter: None
*SAV Save Command
Type: Device status
Description: This command stores the present state of the DC Power Supply and the
states of the current mode in a
specified location in memory.
Syntax: *SAV
Example: *SAV
*SRE Service Request Enable Command/Query
Type: Device status