Programmable DC Power Supply (with Solar Array Simulation) 62000H Series
Operating & Programming Manual
5.6.2.7 IV Subsystem
1. IVCurve:VOC?
Description: Queries the VOC in IV Curve Mode.
Query Syntax: IVCurve:VOC ?
Return Parameter: <NRf+>
Query Example: IVCurve:VOC ?
Return Example: 6.000000e+02
2. IVCurve:ISC?
Description: Queries the ISC in IV Curve Mode.
Query Syntax: IVCurve:ISC ?
Return Parameter: <NRf+>
Query Example: IVCurve:ISC ?
Return Example: 8.000000e+00
3. IVCurve:VMPp?
Description: Queries the VMPp in IV Curve Mode.
Query Syntax: IVCurve:VMPp?
Return Parameter: <NRf+>
Query Example: IVCurve:VMPp?
Return Example: 5.000000e+02
4. IVCurve:IMPp?
Description: Queries the IMPp in IV Curve Mode.
Query Syntax: IVCurve:IMPp?
Return Parameter: <NRf+>
Query Example: IVCurve:IMPp?
Return Example: 5.000000e+00
5. IVCurve:PMPp?
Description: Queries the PMPp in IV Curve Mode.
Query Syntax: IVCurve:PMPp?
Return Parameter: <NRf+>
Query Example: IVCurve:PMPp?
Return Example: 5.000000e+01
6. IVCurve:VT
Description: Creates the 128 dot voltage data table in IV Curve Mode.
Syntax: IVCurve:VT <NRf+>...< NRf+>
Example: Creates the voltage table of IV Curve data.
IVCurve:VT <V1>,<V2,...<V128>
Query Syntax: IVCurve:VT?
Return Parameter: <NRf+>,...,<NRf+>
Query Example: IVCurve:VT?
Return Example: <V1>,<V2>,...<V128>
The execution time of this command is longer in Master/Slave mode. To ensure
normal operation, use the *OPC command. It will run the next command after the
*OPC command responds.
Ex.: IVC:VT xxx,xxx,xxx,...