EasyManua.ls Logo

Chroma A620028 - APG Voltage Calibration

Default Icon
238 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Manual Operation
3-69
set SAVE=YES and press “ ” to save it. If there is no need to save it, press
EXIT
to return to the Calibration screen.
[ S A V E A R G U M E N T ]
S A V E = N O
Figure 3-124
13. Press “
EXIT
” to return to the MAIN PAGE.
Improper shunt range selection may cause damage to the current shunt.
The calibration points may be different for other models (non
62150H-600S); perform the calibration following the instructions
displayed.
3.3.7.4 APG Voltage Calibration
3.3.7.4.1 Hardware Requirements
Table 3-9 lists the hardware requirements for APG voltage calibration.
Table 3-9
Device
Suggest Model or Capacity
DVM
HP 34401A or equivalent DVM
DC Power Supply
Any DC Power Supply or DC signal source
that can output 10Vdc and drive 100mA.
3.3.7.4.2 SETUP
Figure 3-125 shows the wiring connections for APG voltage calibration.
ENTER
CAUTION

Table of Contents

Related product manuals