EasyManua.ls Logo

Chroma A620028 - Current Calibration (62020 H-150 S Only)

Default Icon
238 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Manual Operation
3-77
[
S A V E A R G U M E N T ]
S A V E
=
N O
Figure 3-140
10. Press
EXIT
to return to the MAIN PAGE.
3.3.7.7 IV Current Calibration (62020H-150S Only)
3.3.7.7.1 Hardware Requirements
The hardware required for current calibration is shown in Table 3-12.
Table 3-12
Device
Suggest Model or Capacity
DVM
HP 34401A or equivalent DVM
DCCT
ULTRASTAB SATURN
LOAD
ELECTRICAL LOAD
CHROMA 63204 or equivalent
BREAKER
Current >=100A
The BREAKER capacity listed in the table above is only for the
62150H-600S. For other models, see the OUTPUT CURRENT spec in
Table 1-1 to select the appropriate BREAKER rating.
3.3.7.7.2 SETUP
The wiring diagram for current calibration is shown in Figure 3-141.
62020H- 150 S
+v
v
+
DVM
+sv
sv
Electronic
Load
+
Breaker
ULTRASTAB
SATURN
Figure 3-141
1. Calibrate the offset voltage of the ULTRASTAB SATURN before
using it for the current calibration. Press “
Null
” and the DVM panel
CAUTION

Table of Contents

Related product manuals