EasyManua.ls Logo

Chroma A620028 - Voltage Output & Measurement Calibration (62020 H-150 S Only)

Default Icon
238 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Programmable DC Power Supply (with Solar Array Simulation) 62000H Series
Operating & Programming Manual
3-74
Figure 3-133
7. Press “
” again and the system will set the output voltage of Pin 19 on the rear
panel to 8.0V and the cursor stops at position [4] as shown in Figure 3-133. Input the
voltage read by DVM2 to position [4] and press “ ” to confirm.
8. The APG current calibration is done once the above actions are completed. To save the
calibration data, pressing will display a confirmation page as shown in Figure
3-134. Press ” or turn the “Rotary” ( ) knob to set SAVE=YES and press
” to save it. If there is no need to save it, press “
EXIT
” to return to the Calibration
screen.
[ S A V E A R G U M E N T ]
S A V E =
N O
Figure 3-134
9. Press “
EXIT
” to return to the MAIN PAGE.
The calibration points may be different for other models (non
62150H-600S); perform the calibration following the instructions
displayed.
3.3.7.6 IV Voltage Output & Measurement Calibration (62020H-150S
Only)
3.3.7.6.1 Hardware Requirements
The hardware required for calibration is shown in Table 3-11.
Table 3-11
Device
Suggest Model or Capacity
DVM
HP 34401A or equivalent DVM
[ A P G C U R R E N T C A L I B R A T I O N
]
( M E A . ) O U T P U T V O L T A G E F O R M E A S U R E = 0 . 5 V
A C T U A L
A P G O U T P U T V O L T A G E = 0 . 0 0 0 _ V
( M E A . ) O U T P U T V O L T A G E F O R M E A S U R E = 8 . 0 V
A C T U A L A P G O U T P U T V O L T A G E = 0 . 0 0 0 V
[
3
]
[ 4 ]
ENTER
SAVE
1
ENTER

Table of Contents

Related product manuals