EasyManua.ls Logo

Chroma A620028 - Page 123

Default Icon
238 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Manual Operation
3-75
3.3.7.6.2 Setup
The wiring diagram is shown in Figure 3-135.
Figure 3-135
1. Before perfoming a calibration, ensure the resolution of the test
equipment exceeds the 62020H-150S specifications.
2. Set the Resolution of the HP34401 to SLOW 6 digit.
3. Key in at least 5 digits for each calibration point to ensure the Power
Supply accuracy after calibration.
4. This calibration function is only valid for the 62020H-150S.
3.3.7.6.3 Calibration Procedure
1. Go to the CALIBRATION the screen as shown in Figure 3-136.
[C A L I B R A T I O N]
CHOICE = VOLTAGE [P/M]
1.VOLTAGE [PROG. /MEAS. ]
2.CURRENT [MEAS. ]
3.CURRENT [PROG.]
4.APG VOLTAGE [PROG. /MEAS.]
5.APG CURRENT [PROG. /MEAS. ]
6.IV VOLTAGE[MEAS.] 7.IV CURRENT[MEAS. ]
Figure 3-136
2. In the CALIBRATION page, press or turn the Rotary” ( ) knob to set
CHOICE=1.
3. Press to display the current calibration options as shown in Figure 3-137.
[V O L T A G E C A L I B R A T I O N ]
R E M O V E E X T E R N A L L O A D A N D P R E S S [
E N T E R ] _
S E T T I N G O U T P U T V O L T A G E
1 0 . 0 0 V
A C T U A L O U T P U T V O L T A G E = 0 . 0 0 0 0
V
S E T T I N G O U T P U T V O L T A G E
2 9
. 0 0 V
A C T U A L O U T P U T V O L T A G E = 0 . 0 0 0 0
V
Figure 3-137
1
ENTER

Table of Contents

Related product manuals