EasyManua.ls Logo

Chroma A620028 - Page 124

Default Icon
238 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Programmable DC Power Supply (with Solar Array Simulation) 62000H Series
Operating & Programming Manual
3-76
4. Verify the external load has been removed and press to confirm.
5. The device will output 10.00V and the cursor will stop at position [1] as shown in Figure
3-138. Enter the voltage read by the DVM and press to confirm it.
6. Press “ to do the second low voltage range point calibration. The device will
output 29.00V and the cursor will stop at position [2] as shown in Figure 3-138. Enter the
voltage read by the DVM and press to confirm it.
[V O L T A G E C A L I B R A T I O N]
R E M O V E E X T E R N A L L O A D A N D P R E S S[ E N T E R ]
S E T T I N G O U T P U T V O L T A G E 1 0 . 0 0 V
A C T U A L O U T P U T V O L T A G E= 9 . 9 9 9 8 _ V
S E T T I N G O U T P U T V O L T A G E
2 9 . 0 0
V
A C T U A L O U T P U T V O L T A G E=
2 8 .
V
[ 1 ]
[2 ]
9 9 8 2
Figure 3-138
7. Press “
to do the high voltage range calibration. The device will output 38.00V
and the cursor will stop at position [3] as shown in Figure 3-139. Enter the voltage read
by the DVM and press to confirm it.
8. Press to do the second high voltage range point calibration. The device will
output 132.00V and the cursor will stop at position [4] as shown in Figure 3-139. Enter
the voltage read by the DVM and press to confirm it.
[
V O L T A G E C A L I B R A T I O N
]
R E M O V E E X T E R N A L L O A D A N D P R E S S
[ E N T E R ]
S E T T I N G O U T P U T V O L T A G E 3 8 . 0 0
V
A C T U A L O U T P U T V O L T A G E=
3 7 . 9 9 8 3
V
S E T T I N G O U T P U T V O L T A G E 1 3 2 .
0 0
V
A C T U A L O U T P U T V O L T A G E= 1 3 1 .
V
[
3 ]
[4 ]
9 9 9 1
Figure 3-139
9. The voltage calibration is done when the above actions are completed. To save the
calibrated values, press and a confirmation screen will appear as shown in
Figure 3-140. Press or turn the Rotary” (
) knob to set SAVE=YES and press
to save it. If there is no need to save it, press
EXIT
to return to the Calibration
screen.
ENTER
ENTER
ENTER
ENTER
ENTER
SAVE
1
ENTER

Table of Contents