Programmable DC Power Supply (with Solar Array Simulation) 62000H Series
Operating & Programming Manual
5-28
3. CONFigure:OUTPut
Description: Starts or stops the output voltage/current.
Syntax: CONFigure:OUTPut ON
CONFigure: OUTPut OFF
Parameter: ON|OFF
Example: CONFigure: OUTPut The power supply starts output.
CONFigure: OUTPut OFF The power supply stops output.
Query Syntax: CONFigure: OUTPut?
Query Example: CONF: OUTPut?
Return Example: ON or OFF
4. CONFigure:FOLDback
Description: Sets the type of FOLDBACK PROTECT.
Syntax: CONFigure:FOLDback DISABLE
CONFigure:FOLDback CVTOCC
CONFigure:FOLDback CCTOCV
Parameter: DISABLE|CVTOCC|CCTOCV
Example: CONFigure:FOLD DISABLE
CONFigure:FOLD CVTOCC
Query Syntax: CONFigure:FOLD?
Query Example: CONF:FOLD? Returns the FOLDBACK status.
Return Example: DISABLE or CVTOCC or CCTOCV
5. CONFigure:FOLDT
Description: Sets the delay time for FOLDBACK PROTECT
Syntax: CONFigure:FOLDT <NRf1>
Parameter: 0.01-600.00 (Unit : Sec)
Example: CONF:FOLDT 10
Query Syntax: CONF:FOLDT?
Return Parameter: <NRf1>
Query Example: CONF:FOLDT?
Return Example: 1.000000e+01
6. CONFigure:APGVSet
Description: Sets the APG VSET mode.
Syntax: CONFigure:APGVSet NONE
CONFigure:APGVSet VREF5
CONFigure:APGVSet RREF
Parameter: NONE | VREF5 | VREF10 | IREF | RREF
Example: CONFigure:APGVSet VREF10
Query Syntax: CONFigure:APGVSet?
Query Example: CONFigure:APGVSet?
Return Example: VREF10
7. CONFigure:APGVMeas
Description: Sets the APG VMEAS mode.
Syntax: CONFigure:APGVMeas NONE
CONFigure: APGVMeas VREF5
CONFigure: APGVMeas IREF
Parameter: NONE | VREF5 | VREF10 | IREF
Example: CONFigure:APGVMeas VREF10
Query Syntax: CONFigure: APGVMeas?
Query Example: CONFigure: APGVMeas?
Return Example: VREF10