Reprocessing
124 Instructions for use Oxylog 3000 plus SW 1.n
Removing the disposable breathing
circuit for pediatric patients
1 Disconnect flow measuring lines (A and B).
2 Disconnect the breathing hose (C).
3 Correctly dispose of the complete disposable
breathing circuit. Refer to the chapter
"Disposal" on page 137.
Validated reprocessing procedures
Overview of the reprocessing
procedures for the device and its
components
07637170-2
A
C
B
CAUTION
The disposable breathing circuit must not be
cleaned, disinfected or sterilized: it cannot with-
stand high temperatures and may be damaged.
Device and compo-
nents
Surface disinfection
with cleaning
Machine cleaning with
thermal disinfection
Description of the pro-
cedure
Surface of the device
and other uncritical com-
ponents
Yes No For more information,
refer to "Surface disin-
fection with cleaning"
on page 125.
Breathing valve, breath-
ing hose, flow measuring
lines, angled connector,
flow sensor
No Yes For more information,
refer to "Machine clean-
ing with thermal disinfec-
tion" on page 125.

Table of Contents

Other manuals for Dräger Oxylog 2000 plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 2000 plus and is the answer not in the manual?

Dräger Oxylog 2000 plus Specifications

General IconGeneral
BrandDräger
ModelOxylog 2000 plus
CategoryMedical Equipment
LanguageEnglish

Related product manuals