Instructions for use Oxylog 3000 plus SW 1.n 47
Assembly
Connecting a bacteria filter or HME
The flow and volume measurements can be
corrected for use with an HME. Refer to "Setting the
HME correction" on page 79.
Connect the bacterial filter or HME to the angled
connector as follows.
Adult reusable hose
Adult disposable hose
Disposable pediatric hose
WARNING
Risk of CO2 rebreathing
Bacterial filters, HMEs and, masks increase
the resistance and dead space volume of the
breathing circuit. Note the manufacturers
directions.
NOTE
When using an HME, the measured flow may devi-
ate from the actual expiratory flow, as temperature
and humidity of the gas are reduced.