Technical data
150 Instructions for use Oxylog 3000 plus SW 1.n
Electromagnetic Compatibility (EMC) Complies with IEC 60601-1-2, EN 794-3
(36,101), ISO 10651-3 (36,202,2,1) and UN
Regulation No. 10 (for the power supply unit),
with respect to EMC for use in motor vehicles,
equivalent to European Commission Directive
2004/104/EC.
Complies with RTCA DO-160 with respect to
EMC for use in aircraft and helicopters.
Classification
according to Directive 93/42/EEC
Class IIb
UMDNS code
Universal Medical Device Nomenclature System
18 – 098
Protection class, breathing circuits (disposable or
reusable), including CO
2 sensor,
endotracheal tubes, or masks
Type BF (body floating, defibrillation-proof)
Protection class against ingress of liquids IP34, the device is splash-proof from all directions
and protected against foreign objects >2.5 mm
(0.1 in)
Protection class of the CO
2 sensor IP64
Defibrillation recovery time 0 s
Materials used
NOTE
All Dräger breathing hoses are made without natural
rubber latex.
Reusable breathing circuit for adults
Breathing hose, flow
measuring lines
silicone rubber
Housing of flow sensor,
breathing valve
Polysulfone (PSU)
Vane in flow sensor Stainless steel
Diaphragms in breathing
valve
silicone rubber

Table of Contents

Other manuals for Dräger Oxylog 2000 plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 2000 plus and is the answer not in the manual?

Dräger Oxylog 2000 plus Specifications

General IconGeneral
BrandDräger
ModelOxylog 2000 plus
CategoryMedical Equipment
LanguageEnglish

Related product manuals