Instructions for use Oxylog 3000 plus SW 1.n 125
Reprocessing
Surface disinfection with cleaning
Components:
– Surface of the device and other uncritical
components
Prerequisites:
– The surface disinfectant has been prepared in
accordance with the manufacturer's
instructions.
– The manufacturer's instructions, e.g., regarding
shelf life or application conditions, are
observed.
– An uncontaminated, lint-free cloth soaked in
surface disinfectant is used for the cleaning
surface disinfection.
Cleaning
1 Wipe off obvious soiling using a disposable
cloth (or a similar cloth) soaked in surface
disinfectant and dispose of it.
2 Wipe all surfaces. After that, there must no
longer be any soiling visible.
Surface disinfection
1 Wipe cleaned surfaces again to visibly wet all
surfaces to be disinfected with surface
disinfectant.
2 Wait for the surface disinfectant contact time.
3 At the end of the contact time, moisten a new,
uncontaminated and lint-free cloth with water
(at least drinking water quality).
4 Wipe all surfaces until no remains of the surface
disinfectant, such as foam residue or streaks,
are visible.
5 Wait until the surface is dry.
6 Check the surfaces for visible damage and, if
necessary, replace the product.
Machine cleaning with thermal
disinfection
Use a washer-disinfector that meets the
requirements of the ISO 15883 standard. Dräger
recommends the use of a load carrier for
anesthesia accessories and ventilation
accessories. Follow the manufacturer's instructions
for the washer-disinfector.
Components:
– Breathing valve, breathing hose, flow
measuring lines, angled connector, flow sensor
Surface disinfectant Manufacturer Concentration Contact time
Dismozon plus BODE Chemie 1.6 % 15 minutes
WARNING
Risk due to penetrating liquid
Penetrating liquid may cause the following:
– Damage to the device
– Electric shock
– Device malfunctions
Ensure that no liquid penetrates the device.