Instructions for use Oxylog 3000 plus SW 1.n 145
Technical data
Monitoring
Expiratory minute volume MVe
Alarm, upper alarm limit When the upper alarm limit has been exceeded.
Range of settings 2 to 41 L/min
Alarm, lower alarm limit When the lower alarm limit has been violated.
Range of settings 0.5 to 40 L/min
Airway pressure Paw
Alarm, upper alarm limit When the Pmax value is exceeded.
Range of settings 20 to 60 mbar
Alarm, lower alarm limit When the pressure difference between inspiratory
and expiratory phases is less than 5 mbar for at least
20 seconds.
Or
When the set pressure value is not reached within
10 seconds.
Apnea alarm time Tapn
Alarm When respiratory activity is no longer detected.
Range of settings 15 to 60 s
Apnea
Alarm If no change in respiratory phase has been detected
for >12 seconds and Tapn is deactivated.
Respiratory rate, RRsp
Alarm, upper alarm limit When the upper alarm limit has been exceeded for
more than 25 seconds.
The alarm is silenced for approx. 40 seconds after
starting ventilation or changing the ventilation mode.
Range of settings 10 to 100/min
End-tidal CO2 concentration etCO2 (optional)
Alarm, upper alarm limit When the upper alarm limit has been exceeded for
15 seconds or 3 ventilation cycles.
Range of settings 1 to 100 mmHg/0.1 to 13.3 Vol%/0.1 to 13.3 kPa
Alarm, lower alarm limit When the lower alarm limit has been undercut for
more than 15 seconds or 3 ventilation cycles.
Range of settings 0 to 100 mmHg/0 to 13.3 Vol%/0 to 13.3 kPa