Instructions for use Oxylog 3000 plus SW 1.n 133
Service
Inspection
Inspections must be carried out regularly according
to the following guidelines and within the specified
intervals.
Safety checks
Safety checks are no substitute for preventive
maintenance measures (including preventive
replacement of wear parts) as identified by the
manufacturer.
Scope
1 Check the accompanying documents:
– Instructions for use present.
2 Check components and accessories for
completeness according to the supplied
instructions for use when the product is ready
for operation.
3 Check that the device combination is in
good condition:
– Labels complete and legible.
– No damage.
4 Check electrical safety:
– In accordance with IEC 62353.
5 Check safety functions:
– Correct functioning of the safety valve: max.
pressure 90 mbar.
– Correct functioning of the emergency air
valve.
– Correct functioning of the mains failure
alarm.
– Monitoring of the supply pressure.
– Check the high airway pressure alarm.
– Check the breathing circuit integrity fault
alarm.
– Check the proper functioning of the power
indicators.
6 When operating in a helicopter, make a visual
check of the contacts to the battery and replace
them as necessary.
7 Perform a device check according to the
instructions for use.
Checks Interval Personnel responsible
Inspections and safety checks
1)
Every 2 years Service personnel
1) Designation applies to the Federal Republic of Germany; corresponds to the “Recurring safety inspection” in the Federal
Republic of Austria
WARNING
Risk of medical device failure
If safety checks are not performed on a regu-
lar basis, the proper operation of the medical
device can be compromised.
Perform safety checks at the indicated inter-
vals.