Instructions for use Oxylog 3000 plus SW 1.n 65
Operation
VC-CMV, VC-AC
VC-CMV
Volume Controlled - Controlled Mandatory
Ventilation
Volume-controlled ventilation with a fixed
mandatory minute volume MV set by the tidal
volume VT and the respiratory rate RR. For patients
who are not breathing spontaneously.
Set the ventilation pattern with the controls below
the display:
Tidal volume VT
Respiratory rate RR
(minimum frequency: 5 /min)
Maximum airway pressure Pmax
–O
2 concentration FiO2
The following can be set on the display:
Positive end-expiratory pressure PEEP
Trigger sensitivity
Ventilation time ratio I:E or inspiratory time Ti
When setting the respiratory rate RR, tidal volume
VT or I:E/Ti, the associated values for inspiratory
flow and Ti/I:E are automatically displayed in the
information window.
Plateau time Tplat as a percentage of the
inspiratory time
AutoFlow (optional)
025
Paw
Flow
T
insp
T
e
P
plat
Plateau time
Tplat
1
RR
Insp. Flow
t
t
P
max
PEEP
Exp. Flow
1043717110537171

Table of Contents

Other manuals for Dräger Oxylog 2000 plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 2000 plus and is the answer not in the manual?

Dräger Oxylog 2000 plus Specifications

General IconGeneral
BrandDräger
ModelOxylog 2000 plus
CategoryMedical Equipment
LanguageEnglish

Related product manuals