Instructions for use Oxylog 3000 plus SW 1.n 97
Configuration
Customer Service Mode
In Customer Service Mode, the ventilator performs
function tests, outputs status information and
permits configuration of parameter settings.
The display language in Customer Service Mode is
English. It cannot be changed.
NOTE
Ventilation is not possible when the device is in
Customer Service Mode.
NOTE
The following screens are only intended for main-
tenance personnel and qualified specialists, and
are protected against unauthorized adjustment.
For more information regarding access,
see page 171.
001 Set start-up settings.
(Set start-up settings.)
Configure start-up settings, restore manufacturer's
default settings.
002 Hose/cuvette start-up settings.
(Hose/cuvette start-up settings)
Select which breathing circuit and CO
2 cuvette type to
use.
003 Set date and time.
(Set date and time)
Set date and time.
004
Set measured values display window.
(Set the measured values display window)
Configure the layout of measured values in the measured
values window or restore manufacturer's default settings.
005 Enter activation code.
(Enter activation code)
Enter the activation code for options.
006 Test buttons and potentiometer.
(Test buttons and potentiometer)
Check for correct functioning of keys and controls.
007 Test loudspeaker, buzzer, LEDs and
display. (Test loudspeaker, buzzer,
LEDs and display)
Check for correct functioning of loudspeaker, buzzer,
LEDs and display.
008 Display battery and supply data.
(Display battery and supply data)
Display of battery data and charge status.
009 Check safety valve.
(Check safety valve)
Checking of the safety valve.
010 Display info logbook.
(Display info logbook)
Calibration logbook and technical errors in chronological
sequence.
011 Display user logbook.
(Display user logbook)
Logbook of operating phases, device settings
and alarms.
012 Display maintenance and service
contact information.
(Display maintenance and service
contact information)
Display the maintenance schedule and contact
information for maintenance personnel.
013 CO
2 sensor. (CO2 sensor) Check or calibrate the CO2 sensor.
019 Acoustic alarm configuration Set the minimum alarm volume and alarm tone type.