Instructions for use Oxylog 3000 plus SW 1.n 121
Reprocessing
Semi-critical
Breathing valve, breathing hose, flow
measuring lines, angled connector
Flow sensor
Before reprocessing
Observe before disassembly
1 Turn off the device.
2 Disconnect the plug for the external power
supply.
Patient-specific accessories and consumables
The patient-specific accessories and consumables
must be removed from the device and, if necessary,
disassembled.
Reusable products:
If the reusable product has its own instructions
for use, perform reprocessing in accordance
with the separate instructions for use.
If no separate instructions for use are available
for the reusable product, perform reprocessing
in accordance with the instructions in these
instructions for use.
Disposable products:
Dispose of the disposable products.
Removing the CO
2 sensor and CO2 cuvette
1 Disconnect the CO
2 sensor connector from the
device.
2 Remove the CO
2 sensor (B) from the
CO2 cuvette (C).
3 Remove the CO
2 cuvette (C) from the
flow sensor (D).
4 Remove the angled connector (A)
from the cuvette (C).
052
A
B
C
D

Table of Contents

Other manuals for Dräger Oxylog 2000 plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 2000 plus and is the answer not in the manual?

Dräger Oxylog 2000 plus Specifications

General IconGeneral
BrandDräger
ModelOxylog 2000 plus
CategoryMedical Equipment
LanguageEnglish

Related product manuals