EasyManuals Logo

Dräger Oxylog 2000 plus User Manual

Dräger Oxylog 2000 plus
174 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #75 background imageLoading...
Page #75 background image
Instructions for use Oxylog 3000 plus SW 1.n 75
Operation
Non-invasive ventilation (NIV)
Using NIV
NIV can only be activated as a supplementary
function in the ventilation modes Spn-CPAP (/PS),
PC-BIPAP (/PS), VC-CMV / AF, VC-AC / AF and
VC-SIMV / AF. Oxylog 3000 plus automatically
adjusts to the requirements of non-invasive
ventilation. Mask leakages are detected by the
device and compensated for. Therefore, the
displayed measured values VTe and MVe include
the leakage. The leakage alarm is inactive.
To switch on NIV
1 Select NIV off.
2 Select NIV on and confirm.
The suffix NIV appears in the ventilation mode
field.
WARNING
If NIV is not active, leakage during ventilation
will corrupt the VTe and MVe measured values.
WARNING
Risk of undetected leakage and inadequate
ventilation
Make sure that the NIV mode is not activated
for intubated patients.
WARNING
Risk of CO2 rebreathing
Bacterial filters, HMEs and, masks increase
the resistance and dead space volume of the
breathing circuit. Note the manufacturers
directions.
WARNING
Check MVe alarm limits after deactivating NIV
mode.
WARNING
Risk of aspiration
Avoid high airway pressure.
12837171
WARNING
Risk of inadequate ventilation
The minimum minute volume required by the
patient must be monitored via the lower alarm
limit MVe .
NOTE
Apnea ventilation is not possible when NIV is
active.

Table of Contents

Other manuals for Dräger Oxylog 2000 plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 2000 plus and is the answer not in the manual?

Dräger Oxylog 2000 plus Specifications

General IconGeneral
BrandDräger
ModelOxylog 2000 plus
CategoryMedical Equipment
LanguageEnglish

Related product manuals