Technical data
148 Instructions for use Oxylog 3000 plus SW 1.n
DC/DC converter IP classification IP42
Temperature range –20 °C to +50 °C
Input 12/24/28 VDC; 5 A/2.5 A/2.1 A
Output 19 V ±0.5 V/2.6 A
The intended use of the DC/DC converter is in vehi-
cles.
Gas supply
From a central gas supply system or from an
O2 cylinder.
O
2 supply pressure 270 Pa to 600 kPa at 100 L/min
Supply gas Medical oxygen
Connector for O
2 supply either:
NIST
1)
acc. to EN 739/ISO5359 or
DISS
2)
acc. to CGA V5-1989 or
N-F
3)
S90-116
WARNING
Only use medical grade oxygen.
Gas cylinders and pres-
sure reducers
WARNING
Only use compressed gas cylinders and pres-
sure reducers, which comply with all applica-
ble regulations and have been approved.
Pressure reducer Must have a blow-off valve on the output side to limit
the outlet pressure to approximately 1000 kPa in
the event of a fault.
Gas consumption for
internal control
Average 0.5 L/min
Accuracy of gas
consumption indication
15 % or ±1 L/min, whichever is greater.
1) NIST = Non-Interchangeable Screw-Threaded
2) DISS = Diameter Index Safety Systems
3) N-F = Norme française

Table of Contents

Other manuals for Dräger Oxylog 2000 plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 2000 plus and is the answer not in the manual?

Dräger Oxylog 2000 plus Specifications

General IconGeneral
BrandDräger
ModelOxylog 2000 plus
CategoryMedical Equipment
LanguageEnglish

Related product manuals