Instructions for use Oxylog 3000 plus SW 1.n 57
Getting started
Checking the high airway pressure and
disconnection alarm
Checking the alarm in case of high airway
pressure:
1 Ventilate the test lung in CMV mode.
2 Press the test lung manually, until the airway
pressure exceeds the set Pmax.
3 Check if the Paw high alarm is issued.
Checking the alarm in case of breathing circuit
disconnection:
1 Ventilate the test lung in CMV mode.
2 Disconnect the breathing hose and/or flow
measuring lines from the ventilator.
3 Check if an applicable alarm occurs.
Checking the power supply failure alarm
A monthly check of the power supply failure alarm
is recommended.
1 Switch the device on.
2 Disconnect the external power supply.
3 Remove the battery to activate the acoustic
alarm signal.
4 Listen for the acoustic alarm signal.
5 When the power supply failure alarm check is
completed, re-insert the battery in the battery
compartment of Oxylog 3000 plus.
6 Connect the external power supply.
Troubleshooting
If the device check is not completed successfully:
1 Refer to "Error messages during the device
check" on page 118 of the "Troubleshooting"
section.
2 Contact DrägerService.
NOTE
Contact DrägerService if no alarm is heard.
WARNING
The ventilator is ready for operation only after
all functional tests have been successfully
performed.