Instructions for use Oxylog 3000 plus SW 1.n 93
Monitoring
Checking the CO2 filter during
ventilation
1 Remove the CO2 sensor from the cuvette.
2 Connect the CO
2 sensor (A) to the test filter (B).
3 Press the Settings key.
4 On the screen, select and activate the CO
2
filter check – Run line.
5 Confirm. The filter check starts and the line
displays Busy. During the filter check, the
message ! CO2 meas. inaccurate/failed is
displayed.
During the filter check, ventilation settings can
be changed.
After a successful filter check, the message
Pass is briefly displayed.
6 Attach the CO
2 sensor (A) back to the cuvette.
If the check was not successful:
Oxylog 3000 plus displays the alarm message
!!! CO
2 Filter check failed. The measured
CO
2 value is outside the permissible tolerance
range.
Check whether the sensor (A) or test filter (B) is
soiled and clean them if necessary. Repeat the
CO2 filter check.
177
A
B

Table of Contents

Other manuals for Dräger Oxylog 2000 plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 2000 plus and is the answer not in the manual?

Dräger Oxylog 2000 plus Specifications

General IconGeneral
BrandDräger
ModelOxylog 2000 plus
CategoryMedical Equipment
LanguageEnglish

Related product manuals