Reprocessing
128 Instructions for use Oxylog 3000 plus SW 1.n
Dräger states that oxygen-releasing agents and
chlorine-releasing agents may cause color change
in some materials. Color change does not indicate
that the product is not functioning correctly.
Other surface disinfectants are used at one's own
risk.
Reprocessing procedures
Manual cleaning followed by disinfection by
immersion
The following components can be can be cleaned
manually, followed by disinfection by immersion:
– Breathing valve
– Breathing hose
– Flow measuring lines
– Angled connector
–Flow sensor
The following disinfectants were compatible with
the material at the time of testing:
Steam sterilization
The following components can be steam sterilized:
– Breathing valve
– Breathing hose
– Flow measuring lines
– Angled connector
– Flow sensor
Prerequisites:
– The components have been cleaned,
disinfected, and are dry.
– Use a steam sterilizer that meets the
requirements of the ISO 17665 standard.
Dräger recommends steam sterilization with
fractionated vacuum.
Procedure:
1 Sterilize the components (maximum 134 °C
[273.2 °F], 10 min).
2 Check the components for visible damage and
replace if necessary.
Component Agent Manufacturer
Breathing valve, breathing hose,
angled connector
Cleaning agent:
Neodisher LM2 Dr. Weigert
Disinfectant:
Korsolex Extra BODE Chemie
Flow sensor Cleaning agent:
Sekusept Pulver CLASSIC Ecolab USA
Disinfectant:
Korsolex extra BODE Chemie