Operation
62 Instructions for use Oxylog 3000 plus SW 1.n
Starting operation
The actual screen display may differ in appearance
or configuration.
Switching the device on
To switch the device ON, briefly press the
key (A).
Oxylog 3000 plus performs a self test. The self test
is completed in approximately 6 seconds.
During the self test the start-up screen is briefly
displayed. It includes a progress bar indicating the
progress of the self test, the software version, the
enabled software options, and the prompt to
activate the device check by pressing the rotary
knob (B).
If the rotary knob (B) is not pressed during the self
test, the hose selection page is displayed.
Select the connected hose type by turning the
rotary knob (B) and confirm by pressing the
rotary knob (B).
The ventilator now automatically begins ventilation
with the default settings.
WARNING
Only use a ventilator that has been cleaned,
disinfected and successfully tested to be
ready for operation, to prevent a health risk for
the patient and user.
Refer to the chapter "Reprocessing" on
page 119.
0133717011937171
14937171
NOTE
As long as the hose selection page is shown, the
patient is not being ventilated.
NOTE
The menu to select the hose type can be config-
ured. Refer to "Customer Service Mode" on
page 97.