Technical data
144 Instructions for use Oxylog 3000 plus SW 1.n
Measured values and curves display
Airway pressure measurement
Range 0 to 100 mbar
Resolution 1 mbar
Accuracy ±(2 mbar +2 % of measured value)
Flow measurement
Minute volume MVe
Range 0 to 99 L/min, BTPS
Resolution 0.1 L/min
Accuracy ±15 % of measured value, or ±0.4 L/min, whichever
is greater.
Tidal volume VTe
Range 0 to 5000 mL, BTPS
Resolution 1 mL
Accuracy ±15 % of the measured value or ±20 mL, whichever
is greater (adult breathing hose).
±15 % of the measured value or ±15 mL, whichever
is greater (pediatric breathing hose).
CO2 measurement
Range 0 to 120 mmHg/0 to 15.8 Vol%/0 to 16.0 kPa
Resolution 1 mmHg/0.1 Vol%/0.1 kPa
Total system response
time
200 ms
Frequency measurement
Range 0 to 99/min
Resolution 1/min
Accuracy ±1/min
Curve display
Airway pressure Paw (t) –10 to 100 mbar
Flow (t) –120 to 120 L/min
CO
2 –5 to 120 mmHg/–1 to 16 Vol%/–1 to 16 kPa