Getting started
58 Instructions for use Oxylog 3000 plus SW 1.n
CO2 zero calibration and filter check before ventilation (optional)
The CO2 zero calibration and filter check before
ventilation only work if the CO
2 option has been
installed and the CO2 sensor is present.
1 To switch the device on, briefly press the
key (A).
The device performs a self test and the user is
prompted to activate the configuration menu or
device check:
Press rotary knob for device check and
configuration
2 Press the rotary knob (B) to confirm before the
progress indicator bar is complete.
3 Select CO
2 zero calibration and filter
check in the start-up menu and
confirm.
The function CO
2 zero calibration and filter
check is displayed only if the option is
available.
02211937171
14737171
NOTE
The CO2 zero calibration and filter check can be
discontinued at any time by pressing the
Alarm Reset key.