Instructions for use Oxylog 3000 plus SW 1.n 73
Operation
Apnea ventilation
Apnea ventilation is only applicable when using the
Spn-CPAP mode. In the event of an apnea, the
ventilator will automatically activate volume-
controlled mandatory ventilation (VC-CMV).
When an apnea occurs, the device issues an alarm
signal and, when the alarm time Tapn has been
reached, starts volume-controlled ventilation with
the parameters respiratory rate RRapn, tidal
volume VTapn, and the maximum airway pressure
Pmax. The ventilation time ratio I:E = 1:1.5 and the
plateau time Tplat = 0 are preset during apnea
ventilation.
Setting apnea ventilation
1 Set Tapn with the rotary knob (B) to a value
between 15 and 60 seconds.
The parameters RRapn and VTapn required for
setting apnea ventilation are displayed.
2 Set RRapn and VTapn.
3 Set Pmax to set the maximum permissible
airway pressure during apnea ventilation.
029US
Paw
Flow
Apnea alarm time
Tapn
Start
apnea ventilation
with apnea alarm
t
t
1
RRapn
Pmax
06837170screen 13
Oxylog 3000 plus
A B

Table of Contents

Other manuals for Dräger Oxylog 2000 plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 2000 plus and is the answer not in the manual?

Dräger Oxylog 2000 plus Specifications

General IconGeneral
BrandDräger
ModelOxylog 2000 plus
CategoryMedical Equipment
LanguageEnglish

Related product manuals