Operation
76 Instructions for use Oxylog 3000 plus SW 1.n
Special functions
Manual inspiration / Inspiration hold
The manual inspiration/inspiration hold function will
either initiate a new manual breath or extend the
inspiratory phase of the current breath by up to
15 seconds.
The pattern of the manually started breath
corresponds with the set ventilation mode.
This function is not available for:
– Spn-CPAP without PS
–O
2 inhalation (optional)
Activating Manual inspiration/Inspiration hold
Press the Insp. hold key (A) for the desired
inspiratory time.
100 % O2 (optional)
To apply 100 % O2 for 3 minutes regardless of the
momentarily set value.
Briefly press the 100 % O
2 key (B).
Its indicator lights up for 3 minutes.
After these 3 minutes, or when the 100 % O
2 key is
pressed again, the ventilator continues ventilation
with the currently set O
2 value. The indicator dims.
O2 inhalation (optional)
If a tracheal stenosis or other obstruction occurs,
the flow is interrupted by the ventilator for
5 seconds at an airway pressure of 30 mbar and
the airway pressure is reduced to 0 mbar.
The !!! Paw high alarm is active.
01237170
WARNING
The O2 inhalation function is not a ventilation
mode.
It may only be used for patients with spontane-
ous breathing who receive a constant O2 flow
of between 0 and 15 L/min via a mask.
NOTE
The options 100 % O2 and O2 inhalation are mutu-
ally exclusive.