Instructions for use Oxylog 3000 plus SW 1.n 141
Technical data
Battery
Temperature –20 to +50 °C
(preferred long-term storage temperature <30 °C)
Ambient pressure 570 to 1200 hPa
Relative humidity 5 to 95 % (no condensation)
Settings
Ventilation modes VC-CMV, VC-AC, VC-SIMV, VC-SIMV/PS, Spn-CPAP,
Spn-CPAP/PS, PC-BIPAP, PC-BIPAP/PS
Optional: AutoFlow for VC-CMV, VC-AC and
VC-SIMV(/PS)
Respiratory rate RR 2 to 60/min (VC-SIMV, PC-BIPAP)
5 to 60/min (VC-CMV, VC-AC)
12 to 60/min with apnea ventilation
Ventilation time ratio I:E 1:100 to 50:1
Inspiratory time Ti 0.2 to 10 s
Tidal volume VT 0.05 to 2.0 L, BTPS
Measured values based on the conditions of the
patient's lungs, 37 °C body temperature, airway pres-
sure, water-vapor-saturated gas.
Accuracy of
setting
±15 % of the set value or ±25 mL, whichever is greater
(adult breathing hose).
±15 % of the set value or ±15 mL, whichever is greater
(pediatric breathing hose).
Inspiratory pressure Pinsp PEEP +3 to +55 mbar
O
2 concentration 40 to 100 Vol%
Accuracy of
setting
±10 Vol%. The actual value depends on the inspiratory
flow
1)
and mean airway pressure.
Positive end-expiratory pressure PEEP 0 to 20 mbar, no negative pressure.
Trigger sensitivity (flow trigger) 1 to 15 L/min

Table of Contents

Other manuals for Dräger Oxylog 2000 plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 2000 plus and is the answer not in the manual?

Dräger Oxylog 2000 plus Specifications

General IconGeneral
BrandDräger
ModelOxylog 2000 plus
CategoryMedical Equipment
LanguageEnglish

Related product manuals