Technical data
142 Instructions for use Oxylog 3000 plus SW 1.n
Pressure support ΔPsupp 0 to 35 mbar (relative to PEEP)
Pressure rise time
for pressure support
slow (1 s), standard (0.4 s), fast.
Inspiration
termination of pres-
sure supported
breaths, as a per-
centage of the peak
inspiratory flow (PIF)
(Insp.term.%PIF)
5 to 50 %
Alarm limit ranges
Alarm Alarm limit range
MVe 2 to 41 L/min
MVe 0.5 to 40 L/min
RRsp 10 to 100/min
etCO
2 1 to 100 mmHg/0.1 to 13.3 kPa/0.1 to 13.3 Vol%
etCO
2 0 to 100 mmHg/0 to 13.3 kPa/0 to 13.3 Vol.%
1) see O2 concentration, page 78
Performance characteristics
Control principle Time-cycled, volume-constant, pressure-controlled
Maximum inspiratory flow 100 L/min
1)
Device compliance
with adult breathing
hose,1.5 m
<
1 mL/mbar
with adult breathing
hose, 3 m
<
2 mL/mbar
with pediatric
breathing hose,
1.9 m
<
0.7 mL/mbar
Inspiratory and expiratory resistance of the
device with adult breathing circuit
<6 mbar at 60 L/min
<
4 mbar at 30 L/min
<
2 mbar at 5 L/min