Technical data
142 Instructions for use Oxylog 3000 plus SW 1.n
Pressure support ΔPsupp 0 to 35 mbar (relative to PEEP)
Pressure rise time
for pressure support
slow (1 s), standard (0.4 s), fast.
Inspiration
termination of pres-
sure supported
breaths, as a per-
centage of the peak
inspiratory flow (PIF)
(Insp.term.%PIF)
5 to 50 %
Alarm limit ranges
Alarm Alarm limit range
MVe 2 to 41 L/min
MVe 0.5 to 40 L/min
RRsp 10 to 100/min
etCO
2 1 to 100 mmHg/0.1 to 13.3 kPa/0.1 to 13.3 Vol%
etCO
2 0 to 100 mmHg/0 to 13.3 kPa/0 to 13.3 Vol.%
1) see O2 concentration, page 78
Performance characteristics
Control principle Time-cycled, volume-constant, pressure-controlled
Maximum inspiratory flow 100 L/min
1)
Device compliance
with adult breathing
hose,1.5 m
<
1 mL/mbar
with adult breathing
hose, 3 m
<
2 mL/mbar
with pediatric
breathing hose,
1.9 m
<
0.7 mL/mbar
Inspiratory and expiratory resistance of the
device with adult breathing circuit
<6 mbar at 60 L/min
<
4 mbar at 30 L/min
<
2 mbar at 5 L/min

Table of Contents

Other manuals for Dräger Oxylog 2000 plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 2000 plus and is the answer not in the manual?

Dräger Oxylog 2000 plus Specifications

General IconGeneral
BrandDräger
ModelOxylog 2000 plus
CategoryMedical Equipment
LanguageEnglish

Related product manuals