Instructions for use Oxylog 3000 plus SW 1.n 151
Technical data
EMC declaration
General Information
The EMC conformity of Oxylog 3000 plus includes
the use of the following external cables,
transducers and accessories:
AC/DC power pack
DC/DC converter
Multi-purpose wall holder
Equipment holder
Carrying bag
Quick Power Connector
Carrying System
–CO
2 sensor
–CO
2 extension cable
Data communication cable
This device was tested for electromagnetic
compatibility using accessories from the list of
accessories. Other accessories may only be used
if they do not compromise the electromagnetic
compatibility. The use of non-compliant
accessories may result in increased
electromagnetic emissions or decreased
electromagnetic immunity of the device.
This device may be used in the direct vicinity of
other devices only if Dräger has approved this
device arrangement. If no approval has been given
by Dräger, it must be ensured that this device
functions correctly in the desired arrangement
before use. The instructions for use for the other
devices must be followed.

Table of Contents

Other manuals for Dräger Oxylog 2000 plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 2000 plus and is the answer not in the manual?

Dräger Oxylog 2000 plus Specifications

General IconGeneral
BrandDräger
ModelOxylog 2000 plus
CategoryMedical Equipment
LanguageEnglish

Related product manuals