Instructions for use Oxylog 3000 plus SW 1.n 67
Operation
Deactivating the trigger
1 Set a value less than 1 L/min or greater than
15 L/min (off is displayed instead of a value).
2 Press the rotary knob to confirm.
The last effective trigger value is adopted by the
ventilator when changing from VC-AC to PC-BIPAP
or SpnCPAP.
Activating AutoFlow (optional)
The AutoFlow-AF function can be set for VC-CMV
and VC-AC.
When AutoFlow is switched on, the setting Tplat is
no longer valid, and Slope must be set.
For more information on AutoFlow, refer to
"AutoFlow" on page 158.
Cardiopulmonary resuscitation (CPR)
During cardiopulmonary resuscitation, the airway
pressure Paw is increased because of chest
compressions.
Oxylog 3000 plus will try to limit the airway
pressure Paw to the set Pmax without ending the
inspiration prematurely.
However, if, due to compressions, the airway
pressure Paw exceeds the set Pmax by 5 mbar,
Oxylog 3000 plus cycles to the expiratory phase.
Therefore in general, if Pmax is set to a higher
value, a higher minute volume is possible.
However, this increases the intrathoracic pressure
and may reduce coronary perfusion.
10737171-1

Table of Contents

Other manuals for Dräger Oxylog 2000 plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 2000 plus and is the answer not in the manual?

Dräger Oxylog 2000 plus Specifications

General IconGeneral
BrandDräger
ModelOxylog 2000 plus
CategoryMedical Equipment
LanguageEnglish

Related product manuals