Operation
74 Instructions for use Oxylog 3000 plus SW 1.n
Ending apnea ventilation
Press the Alarm Reset key.
The ventilator is again ventilating with the previous
settings in ventilation mode Spn-CPAP.
Deactivating apnea ventilation
Set Tapn to OFF.
Spn-CPAP/PS
If
Δ
Psupp is set to a value above 0 mbar, the
following parameters can be adjusted on-screen in
addition to those for Spn-CPAP:
Trigger sensitivity
Successful triggering by the patient is indicated
by an asterisk (*) on the left side of the curve
window.
Pressure rise time Slope (effective for the
pressure support
Δ
Psupp)
Cardiopulmonary resuscitation (CPR)
Refer to "Cardiopulmonary resuscitation (CPR)" on
page 67.
NOTE
Apnea ventilation can only be activated in ventila-
tion mode Spn-CPAP without NIV.
The minimum minute volume required by the
patient must be monitored via the lower alarm limit
MVe .